CONTACT US
Product Overview
The combustion efficiency analyzer utilizes Non-Dispersive Infrared (NDIR) technology and long-life electrochemical sensors to measure the volume concentrations of CO, CO₂, and O₂ in flue gas in real time.
By combining flue gas temperature and excess air coefficient calculations, the analyzer provides a comprehensive evaluation of combustion performance. It is ideal for optimizing industrial combustion processes, improving energy efficiency, reducing fuel consumption, and supporting energy-saving and emission reduction goals.
Working Principle
The analyzer uses NDIR technology to measure CO and CO₂ concentrations based on the absorption of specific infrared wavelengths.
Electrochemical sensors are used for O₂ detection. By integrating flue gas temperature data and excess air coefficient models, the system performs real-time calculations to assess and optimize combustion efficiency.
Key Features
- Multi-Parameter Measurement: Simultaneously measures CO, CO₂, and O₂ concentrations, and calculates flue gas temperature and excess air coefficient.
- High Accuracy: Precision sensor design with wide measurement range prevents over-limit damage and ensures long-term stability.
- Portable Design: Integrated sampling system with compact structure, suitable for both field and laboratory use.
- Smart Power Management: Built-in intelligent battery management and soft-start function extend service life.
- Convenient Data Management: Supports automatic data storage, retrieval, deletion, and transmission via multiple interfaces.
Applications
- Industrial Combustion Processes: Used in boilers, furnaces, and thermal equipment to optimize combustion and reduce energy consumption.
- Petrochemical, Power, and Chemical Industries: Enables real-time monitoring to improve production efficiency and reduce emissions.
- Energy Saving & Emission Reduction: Optimizes combustion through excess air coefficient analysis, lowering greenhouse gas emissions.
- Laboratory & Field Testing: Suitable for combustion research and on-site measurements with fast response and easy operation.
| Parameter | Description |
| Measured Components | CO, CO2, O2, Flue Gas Temperature; Optional SO2, NO, NO2 sensors |
| Measurement Principle | O2: ECD; CO/CO2: NDIR; Flue Gas Temperature: K-type thermocouple or S-type thermocouple |
| Measurement Range | O2:(0~21)%;CO:(0~5000)ppm |
| CO2:(0~25)%;Flue Gas Temperature:(0~800~1200)℃ | |
| Linear Error | CO/CO2: ±1%F.S.; O2: ±2%F.S.; Flue Gas Temperature: ±3℃ or measured value ±1% |
| Resolution | O2: 0.01%; CO: 1ppm; CO2: 0.01%; Flue Gas Temperature: 0.1℃ |
| Repeatability | ≤2% |
| Operating Temperature | (0~50)℃ |
| Relative Humidity | (5~85)% |
| Response Time | T90<15s (NDIR) |
| Communication | RS-232/RS-485 |
| Battery Power | Built-in rechargeable lithium battery, external 12.6V charger, can work continuously for more than 10 hours |
| AC Power Supply | 100~240VAC, 80W |
| Display | LCD Display |
| Features | Built-in imported sampling pump, pressure-resistant capability of 200hPa |
| Self-diagnosis function, online sensor status check | |
| Optional pre-treatment device |
Wobo Industrial Equipment Co., Ltd.