The Laser Trace Moisture Gas Analyzer uses Tunable Diode Laser Absorption Spectroscopy (TDLAS) technology and is specially designed for precise gas moisture measurement. It provides real-time readings of moisture volume fraction and dew point. The analyzer offers high precision, fast response, high sensitivity, and long-term stability, making it ideal for natural gas liquids (NGL) recovery units, low-temperature LNG processes, and other industrial gas applications. The device supports multiple signal outputs and communication interfaces, and features explosion-proof construction and high protection ratings for safe operation in harsh industrial environments.
Technical Principles
The instrument employs TDLAS technology to measure the absorption of laser light at specific wavelengths by water molecules in the gas. This enables high-precision, online measurement of gas moisture and dew point with minimal interference from background gases. The fast response time and long-term stability ensure reliable monitoring under continuous operation.
Key Features
- Customizable Measurement Range: Flexible range settings tailored to process requirements.
- TDLAS Technology: High sensitivity, fast response, and long sensor lifespan.
- High Precision: Accurate and reliable moisture and dew point readings.
- Stable Operation: Continuous operation without interference from background gases.
- No Cooling Required: Simple operation without additional cooling equipment.
- Multiple Signal Outputs: Supports analog and digital outputs, compatible with various control systems.
- Explosion-Proof Design: Suitable for flammable and explosive environments.
- High Protection Rating: Meets stringent industrial environmental requirements.
Applications
- Natural Gas Liquids (NGL) Recovery Units: Monitors moisture and dew point in low-temperature processes to ensure process safety and efficiency.
- Liquefied Natural Gas (LNG) Production: Accurate measurement of gas moisture and dew point, preventing icing or corrosion of equipment.
- Natural Gas Pipeline Transport: Monitors moisture content to prevent hydrate formation, ensuring safe transport.
- Industrial Gas Drying: Provides precise humidity control, ensuring high-quality drying processes.
- Environmental Monitoring: Measures gas moisture to guarantee reliable and accurate monitoring results.
| Parameter | Description |
| Measurement Component | H2O |
| Measurement Principle | TDLAS |
| Measurement Range | 0~50ppm~6000ppm |
| Measurement Accuracy | ±2%F.S. |
| Resolution | 0.1 ppm |
| Repeatability | ≤1.5%F.S. |
| Explosion-proof Rating | Ex d IIC T6 |
| Protection Rating | IP66 |
| Ambient Temperature | -20℃~60℃ |
| Operating Humidity | 0~98%RH (non-condensing) |
| Operating Pressure | Inlet: 140~310kPa; Gas Cell: 800~1200mbara |
| Power Supply | 24 VDC±20% or 100~240 VAC±10%, 50/60Hz, 10W |
| Communication Interface | RS-485, 4~20mA, Modbus, Ethernet |