Search productSearch post

Portable Gas Analyzer for Coal Gas
Portable Gas Analyzer for Coal Gas
Portable Gas Analyzer for Coal Gas
Portable Gas Analyzer for Coal Gas

Portable Gas Analyzer for Coal Gas

  • Keywords:portable gas analyzer, coal gas analyzer, gas concentration measurement
  • Category:CO2 Analysis and Detection

The Portable Gas Analyzer for Coal Gas is a high‑precision instrument designed for rapid, on‑site measurement of key components in coal gas, including CO, CO₂, CH₄, H₂, O₂, and other combustible or toxic gases. Using advanced infrared, electrochemical, or thermal conductivity sensing technologies, it delivers accurate and stable readings even in harsh industrial environments. Its compact and rugged design makes it ideal for field inspections, coal gas production monitoring, pipeline safety checks, and combustion optimization. With fast response time, intuitive operation, and long‑term reliability, this analyzer ensures efficient and safe coal gas analysis across industrial, energy, and environmental applications.

CLICK TO INQUIRE

Product Overview

The Portable Gas Analyzer (Coal Gas) integrates Non-Dispersive Infrared (NDIR), long-life electrochemical sensors, and MEMS thermal conductivity technology to simultaneously measure the calorific value of coal gas and biogas while analyzing the volume concentrations of CO, CO₂, CH₄, H₂, O₂, CₙHₘ, and other gases. Its portable design allows for industrial field measurements and laboratory gas sampling, efficiently replacing traditional combustion calorimeters. The analyzer is ideal for coal gas and natural gas calorific value measurement and multi-component gas analysis.

Technical Principles

  • NDIR Technology: Measures CO, CO₂, CH₄, and other infrared-absorbing gases.
  • Electrochemical Sensors: Measures O₂, H₂, and other reactive gases with high precision.
  • MEMS Thermal Conductivity: Determines the calorific value of CₙHₘ gases.
  • This combination allows simultaneous multi-component gas measurementand accurate calorific value calculation without mutual interference.

 

Key Features

  • Multi-Gas and Calorific Value Measurement: Measures CO, CO₂, CH₄, H₂, O₂, CₙHₘ simultaneously and calculates coal and natural gas calorific value automatically.
  • High Measurement Accuracy: Each gas component does not interfere with the measurement of others. For example, CₙHₘ does not affect CH₄ measurement; CO, CO₂, CH₄ do not interfere with H₂ measurement.
  • Portable Design: Lightweight and compact, ideal for field operation and laboratory gas bag sampling.
  • Smart Power Management: Built-in battery management with soft-start power control, extending battery life.
  • Efficient Data Management: Automatic data storage, with options for query, deletion, and multiple data interfaces.
  • Low Operating Cost: No disassembly, chemical reagents, or consumables required; simple, cost-effective operation.

 

Applications

  • Coal Gas and Natural Gas Monitoring: Field measurement and calorific value analysis of coal gas, natural gas, and biogas, including coke oven gas and mixed coal gas.
  • Industrial Sites: Monitoring combustion processes in power, petrochemical, and chemical industries to improve fuel efficiency and save energy.
  • Laboratory Analysis: Accurate sampling and analysis of coal and natural gas compositions for research and quality control.
  • Environmental Monitoring: Measures CO, CO₂, H₂, and other emissions to ensure compliance with environmental standards.
  • Energy Management: Real-time tracking of fuel calorific value to optimize energy consumption and reduce costs.
Parameter Description
Measured Components CO/CO2/CH4/H2/O2/C2H2/C2H4/CnHm*/Heating Value
Measurement Principle CO/CO2/CH4/C2H2/C2H4/CnHm: NDIR; H2: TCD; O2: ECD
Measurement Range CO/CO2/CH4/H2:(0~100)% O2:(0~25)%;
C2H2/C2H4/CnHm: (0~10)% (optional)
Linear Error CO/CO2/CH4/C2H2/C2H4/CnHm: ±1%F.S.; O2/H2: ±2%F.S.
Resolution 0.01%
Repeatability ≤1%
Response Time T90<15s (NDIR)
Optimal Flow (0.7~1.2)L/min
Inlet Pressure (2~50)kPa
Sample Gas Requirement Dust-free, water-free, oil-free
Dimensions (L*W*H) 412*152*294(mm)
Weight 9kg
Communication RS-232
Power Supply Built-in rechargeable lithium battery, external 12.6V charger
Display LCD Display
Features Built-in sampling pump
Optional pre-treatment device
Heating value units Kcal/m3 and Mj/m3 can be switched freely
Self-diagnosis function, online sensor status check
Note CnHm refers to the total of hydrocarbon gases other than methane.
Expand more!
Portable Gas Analyzer for Coal GasPortable Gas Analyzer for Coal Gas

Click on the inquiry